Automatic Interpretation of RPR Tests Using Lightweight Hybrid Architectures for Binary and Ternary Classification: A Preliminary, Single-Device Proof-of-Concept Study
Data
2026
Embargo
Autores
Orientador
Título da revista
ISSN da revista
Título do volume
Editora
Título Alternativo
Resumo
This study evaluates a lightweight, edge-deployable artificial intelligence pipeline to assist,
not replace, trained human readers in the classification of RPR test reactions. Two separate
and non-directly comparable experimental configurations were investigated: a binary
task (Reactive vs. Non-Reactive) using 243 original images and a ternary task (Reactive,
Minimally Reactive, Non-Reactive) using a distinct dataset of 293 original images. Because
the datasets were acquired using a single device and laboratory protocol, and because
deterministic augmentation generates highly correlated transformations rather than inde
pendent clinical samples, the reported results should be interpreted as preliminary internal
evidence of feasibility rather than proof of clinical generalizability. In the augmented
internal test evaluation, the binary model achieved 99.98% accuracy (25,137/25,200), while
the ternary model achieved 91.12% accuracy (14,417/15,822). In the original-image deploy
ment evaluation, binary performance remained 100% (58/58) across FP32, FP16, and INT8;
ternary performance was preserved under FP32/FP16 at 95.24% (80/84) but decreased to
76.19% (64/84) after INT8 quantization. An additional stochastic augmentation experiment
for ternary INT8 deployment restored performance to 95.24% (80/84) and 0.9444 Macro-F1,
but external validation remains mandatory before any clinical adoption.
Palavras-chave
RPR test, syphilis screening, deep learning, edge AI, lightweight models, medical image classification, ONNX
Tipo de Documento
Artigo
Versão da Editora
Dataset
Citação
Identificadores
TID
Designação
Tipo de Acesso
Acesso Aberto