Characterization of Ultrasound Probe-Dependent Interference in Electromagnetic Tracking for Image-Guided Procedures
| dc.contributor.author | Valente, Simão | |
| dc.contributor.author | Morais, Pedro | |
| dc.contributor.author | Fritz, Andreas | |
| dc.contributor.author | Stern, Antonia | |
| dc.contributor.author | Lima, Estevão | |
| dc.contributor.author | Vilaça, João | |
| dc.date.accessioned | 2026-07-15T10:58:36Z | |
| dc.date.available | 2026-07-15T10:58:36Z | |
| dc.date.issued | 2026 | |
| dc.description.abstract | Ultrasound (US) imaging is widely used to guide minimally invasive procedures such as percutaneous nephrolithotomy (PCNL), while electromagnetic (EM) tracking can complement US guidance by providing line-of-sight-independent instrument localization. However, US probes may distort the EM tracking field in a probe-dependent manner. This study characterized probe-induced EM interference for a conventional 3D/4D phased-array probe and a handheld wireless probe. Three experiments were conducted using an EM tracking system: spatial mapping of interference along each probe body, assessment of probe–sensor separation for the handheld probe, and evaluation of probe-induced tracking deviations in a simulated EM-guided PCNL setup with tracked needle and catheter sensors. EM-US calibration was then performed using low-interference sensor positions. The phased-array probe produced minimal disturbance, maintaining submillimetric positional and subdegree orientational precision across tested modes. Compared with the evaluated phased-array probe, the evaluated handheld wireless probe generated stronger, spatially localized interference, requiring ≥75 mm positional and ≥50 mm orientational separation to recover baseline precision. In the PCNL simulation, the phased-array probe maintained tracking stability, whereas the handheld probe introduced localized deviations. Both probes produced RMS calibration residuals below 1 mm under controlled conditions. These results provide device-specific baseline measurements and a workflow for probe-dependent interference assessment and sensor-placement optimization in EM-US navigation. | |
| dc.identifier.uri | https://hdl.handle.net/11110/3427 | |
| dc.rights | open access | |
| dc.rights.uri | http://creativecommons.org/licenses/by/4.0/ | |
| dc.subject | electromagnetic tracking | |
| dc.subject | image-guided interventions | |
| dc.subject | ultrasound technologies | |
| dc.subject | minimally invasive surgery | |
| dc.title | Characterization of Ultrasound Probe-Dependent Interference in Electromagnetic Tracking for Image-Guided Procedures |